6.4 - Equipements
6.4.1 The laboratory shall have access to the equipment required for the correct performance of laboratory activities. This includes, but is not limited to, measuring instruments, software, measurement standards, reference materials, reference data, reagents and consumables, auxiliary apparatus, or any combination thereof, necessary for laboratory activities and capable of influencing the result.
6.4.2 In cases where the laboratory uses equipment that it does not permanently control, it shall ensure that the equipment-related requirements of this International Standard are met.
6.4.3 The laboratory shall have a procedure for the handling, transport, storage, use, and planned maintenance of equipment in order to ensure proper functioning and to prevent contamination or deterioration.
6.4.4 The laboratory shall verify that equipment conforms to specified requirements before it is placed into service.
6.4.5 Measuring instruments shall be capable of achieving the measurement accuracy and measurement uncertainty required to provide a valid result.
6.4.6 Measuring equipment shall be calibrated when:
the measurement accuracy or measurement uncertainty affects the validity
of the reported results; or
calibration of the equipment is required to establish the metrological
traceability of the reported result.
6.4.7 The laboratory shall establish a calibration program, which shall be reviewed and adjusted as necessary in order to maintain confidence in the calibration status.
6.4.8 Any equipment requiring calibration, or for which a period of validity has been established, shall be labeled, coded, or otherwise identified to allow the user of the equipment to readily identify the calibration status or period of validity.
6.4.9 Equipment that has been subjected to overloading or mishandling, gives questionable results, or has been shown to be defective or outside specified requirements, shall be taken out of service. It shall be isolated to prevent its use, or clearly labeled or marked as being out of service, until it has been repaired and its correct functioning has been verified. The laboratory shall examine the effect of the defect or deviation from specified requirements and shall initiate the procedure for the management of nonconforming work (see 7.10).
6.4.10 When intermediate checks are needed to maintain confidence in the performance of equipment, they shall be carried out according to a procedure.
6.4.11 When calibration data and reference material data include reference values or correction factors, the laboratory shall ensure that these reference values and correction factors are updated and implemented, as appropriate, in order to meet specified requirements.
6.4.12 The laboratory shall ensure that practicable measures are taken to prevent unintended adjustments of equipment that would invalidate the results.
6.4.13 Records shall be established for equipment that can influence
laboratory activities. The records shall include at least the following:
a) the identity of the equipment, including software and its version
and/or firmware and its version;
b) the manufacturer’s name, type identification, and serial number or
other unique identification;
c) evidence of verification that the equipment conforms to specified
requirements;
d) the current location;
e) the dates of calibration, the results of all calibrations, adjustments,
acceptance criteria, and the due date of the next calibration or the
calibration interval;
f) the dates, results, and documentation of reference materials, the
acceptance criteria, and the period of validity;
g) the maintenance plan and up-to-date information on maintenance performed,
where relevant to the performance of the equipment;
h) details of any damage, malfunction, modification, or repair of the
equipment.
![]()
The laboratory must be equipped with suitable equipment for the activities performed, with reference to all phases and all aspects of these activities, including sample management, preparation of the items tested, analysis, and data processing.
If the laboratory uses equipment provided by a third party, or in any case equipment outside its permanent control, it must ensure that such equipment meets the applicable requirements. The standard does not impose any requirement regarding ownership or property.
Equipment, including software, must enable the required accuracy to be achieved.
Equipment must comply with the applicable technical specifications and must be checked before being placed into service and during use through appropriate periodic verifications or calibrations.
The conformity assessment body must provide evidence of the metrological conformity of its equipment.
Metrological confirmation is an operation performed by the laboratory using the measuring equipment.
Equipment must be managed according to criteria of technical and economic effectiveness.
Personnel must have all the information and knowledge necessary to ensure the best possible use of the equipment, in relation to their assigned duties.
Equipment, including software and firmware, must be clearly and uniquely identified.
Information relating to the characteristics and operation of test and measuring equipment must be available in writing, either as descriptions prepared by the laboratory or as operating and handling manuals provided by the manufacturer.
In this case, the laboratory must also document the full availability and accessibility of the corresponding documentation and operating instructions.
As a general rule, equipment must be used only by strictly authorized personnel.
If, for any reason, a piece of equipment falls outside the laboratory’s control, the laboratory must ensure that it is functioning properly before returning it to service.
If, following calibration, the use of correction factors becomes necessary, these factors must be recorded, and the continued applicability of those factors must be verified and confirmed.
The use of correction factors is not recommended, except in cases of absolute necessity.
The laboratory must maintain a list of all equipment, including the associated software, that has an influence on the tests performed.
Each piece of equipment must be accompanied by a set of records including:
- The name of the equipment and its internal identification code;
- The manufacturer’s name, model, serial number, or any other form of
unique identification;
- The date of delivery and the date it was put into service;
- Verification of conformity with specifications, for example the incoming
inspection report;
- Its current location;
- Reference to the descriptive documentation, such as operating instructions,
manuals, etc.;
- Dates, results, reports, and certificates relating to verification,
calibration, checks, adjustments, etc. performed on the equipment, as
well as the dates of the next scheduled repetitions of these operations;
- Maintenance programs and records of maintenance performed;
- Records of damage, repairs carried out, and modifications made.
The equipment must be protected against deterioration, damage, misuse, and alteration, and must be subject to regular maintenance.
Equipment that has been subject to failure, overload, misuse, or that appears to be defective or nonconforming must be identified and isolated until it has been restored and its conformity has been verified.
The laboratory must assess whether such failures may have had any impact on previously performed activities and take appropriate action accordingly.
Where applicable, equipment may bear a label indicating its “calibration or verification status,” with the date of completion and the due date or the date of the next verification. However, the standard does not require this information to be physically displayed on the equipment.
If intermediate checks are necessary to ensure confidence in the calibration status, they shall be performed according to predefined procedures.
The criteria presented apply equally to simple analog devices and computerized systems.
The automation of testing processes is an important factor in improving the quality of laboratory services.
It leads to improved efficiency and productivity, while reducing the risk of human error.
In addition, it frees personnel from certain routine tasks and allows them to focus on more “creative” activities, thereby supporting their professional growth.
The use of computerized systems for managing testing processes should therefore be an objective for all laboratories, and the quality management of these systems represents a benchmark for the quality of a modern laboratory.
Remember to specify any applicable restrictions, for example: a tensile testing machine verified in tension but not in compression, and up to 20 metric tons, even though the load cell has a capacity of 25 metric tons.
Certain more sensitive pieces of equipment must undergo daily maintenance to ensure optimal operation and results consistent with the targeted measurement uncertainty. The protocol must be covered by a maintenance plan, including management of consumables, recorded modifications, verification of proper equipment operation after maintenance, criteria for taking the equipment out of service or requiring mandatory repair, and a restart procedure after an extended period of inactivity.
The laboratory shall select and use reference materials that are suitable for their intended use in the measurement process.
The standard requires intermediate checks to be performed when necessary. If the CAB uses reference materials, they must be managed like any other “standard” with a limited period of validity: related to their consumption, the guarantee of their intrinsic properties, and indirectly to their packaging or storage conditions, which may require specific arrangements such as controlled temperature, controlled humidity, preservation, etc.
The dates, results, and documentation related to reference materials, as well as the acceptance criteria and period of validity, shall be formally documented.