Bias and Measurement Uncertainty
Many instruments provide results with a bias, meaning a deviation from the true value, which can either overestimate or underestimate the value being measured.
This systematic drift is sometimes difficult to detect, especially when it is small compared to the result. In such cases, the operator generally does not take it into account. However, in trace analysis, for example, its impact may have significant consequences, and it will therefore need to be included either in the result or in the uncertainty.
It is this second option, uncertainty, that we will try to clarify.
- The final uncertainty shall be greater than or equal to the uncertainty
that would have been reported if the bias had been corrected.
- Underestimating the uncertainty indicates an invalid uncertainty statement;
similarly, overestimating the uncertainty indicates a poorly constructed uncertainty
budget.
- The method shall be limited to the method proposed by the Guide (GUM) when
the bias correction is applied.
- For any coverage factor and any amplification of the bias, the confidence
level of the expanded uncertainty should be at least the level obtained when
the bias is corrected. For example, if the distribution of values that could
reasonably be attributed to the measurand is Gaussian, then k = 2 should imply
at least 95% confidence.
- The method shall be transferable so that the uncertainty and bias of a result
can be used as components in another uncertainty statement.
Measurement results shall be corrected for bias, and the uncertainty associated
with the bias correction should be included as a contribution to the combined
standard uncertainty.
However, when correction for measurement bias is not possible, the bias should
be explicitly taken into account in the uncertainty statement.
A complete uncertainty statement includes the combined standard uncertainty,
calculated as if the measurement result had been corrected for bias; an explicit
statement of the bias value; and an expanded uncertainty that includes a term
accounting for the bias.
The usual method of using the expanded uncertainty U, for a measurement result
y that has an unknown “true†value of the measurand Y, is to produce an uncertainty
interval, with a confidence level determined by the coverage factor k, such
that:
![]()
Therefore, the measurement result is often expressed as:y ± U.
In cases where the result is corrected for bias δ, uA similar uncertainty interval can be constructed for the corrected measurement result y
y
This is equivalent to the uncertainty interval of: Y– (U + δ)≤Y≤Y + (U – δ)
Accordingly, the measurement result could be stated as:

Unfortunately, this may lead to the possibility that one of the uncertainty bounds could become negative, for example, if the bias is positive and δ>U then the upper limit of the uncertainty will be negative.
This may create confusion for operators, particularly when making a statement of conformity. Therefore, an additional condition must be imposed: the uncertainty limits shall be greater than or equal to zero for all values of δ.

If we compare the proposed method for handling uncorrected bias with two other techniques that have been proposed to address this issue:
The first method treats the uncorrected bias as another source of uncertainty and simply combines it with the combined standard uncertainty. This method is denoted RSSuc.
The second method adds the bias to the expanded uncertainty. This method is denoted RSSU.
In contrast, the third proposed method combines the bias algebraically with the expanded uncertainty, except when the bias is large. This method is referred to as SUMU.
The three methods are presented below:
- Method RSSuc : Y = y ± U (RSSuc) with:
Often used consensually in standards, but not statistically relevant
- Method RSSU : Y = y ± U (RSSU) with:

- Method SUMU : Y = y ± U with:
![]()
with δ>0 and if k.u
![]()
withδ<0 and if k.u
Significant differences can be observed between the three methods.
The three graphs below show the actual statistical confidence of the three methods as a function of the uncorrected bias for coverage factors k = 1, 2, and 3.
The distributions are assumed to be Gaussian (normal) in all cases.

For example, in the case where k = 2, we would ideally like to maintain 95% confidence for all values of the uncorrected bias.
The SUMU method maintains this level of confidence until the ratio of the bias to the combined standard uncertainty becomes greater than the coverage factor.
For large bias values, the SUMU method produces uncertainty intervals that are slightly more conservative, meaning larger than necessary to provide valid 95% confidence levels.
The RSSuc method, on the other hand, may produce uncertainties that are much larger than necessary.
For example, with k = 2 and a bias twice as large as the combined standard
uncertainty (δ/u
Although this overestimation of uncertainty is not necessarily catastrophic, it may lead to a greater reduction in the specification zone.
Conversely, the RSSU method significantly underestimates the true uncertainty.
For example, with a coverage factor k = 2 and an uncorrected bias twice as
large as the combined standard uncertainty (δ/u

The scale on the left-hand y-axis defines the full width of the uncertainty interval divided by the combined standard uncertainty. Therefore, the ratio would be equal to 2k, where k is the coverage factor, if the bias had been corrected.
As can be seen, the SUMU method always produces the smallest expanded uncertainty interval compared with the other methods, for all bias values and coverage factors.
An example of how the size of the expanded uncertainty could have an impact is shown on the right-hand y-axis.
This axis describes the percentage of the specification zone that is used by the expanded uncertainty interval for the rather atypical case where the ratio of the specification zone to the expanded uncertainty interval, if the bias had been corrected, is 4:1.
The SUMU method uses the smallest percentage of the specification zone compared with the other two methods.
For example, for k = 2 and an uncorrected bias equal to four times the combined
standard uncertainty (δ/u
The figures illustrate that, of the three methods described, the SUMU method provides a significant advantage when the uncorrected bias is included in the uncertainty.
This method always maintains an actual confidence level higher than the nominal confidence level corresponding to the coverage factor used.
Although these examples are based on a Gaussian distribution, the SUMU method maintains this relationship between k and the confidence level for any type of distribution, because the resulting interval always contains at least the same interval as if the result had been corrected.
The SUMU method minimizes the percentage of the specification range used by the expanded uncertainty interval.
In addition, this method avoids negative expanded uncertainties, which could be a source of confusion for the operator when determining the conformity zone.
The following examples are established using the consensus method: RSSuc.
![]()
Consider a measurement result, Y, that has a bias δ
Assuming that δ
Let s denote the experimental standard deviation of the n measurements; the estimated Type A uncertainty is then s/√n.
u
In this example δ1<0 and ku

Suppose that a measuring device is designed to inspect parts with a length
of 100 mm and has been verified using a reference gauge block with
a combined standard uncertainty of U
5 measurements were recorded, with an experimental standard deviation of 3.0 µm and a mean result 4.0 µm below the reference value: the bias is negative.
Within-laboratory uncertainty: u
Therefore:


with a 95% level of confidence (K=2)
![]()
For certain measurements, the bias may be estimated rather than directly measured. For example, length measurements are often not corrected to the standard temperature of 20°C.
Therefore, the uncorrected thermal expansion represents a measurement bias. Suppose the temperature of a metal bar varies between 20°C and 30°C, with an estimated average of 25°C.
The estimated magnitude of the bias is given by δ2 (δ2>
0) which represents the length deviation due to uncorrected thermal
expansion over the 5°C temperature difference.
Temperature variability can be described by a rectangular distribution over
10°C, i.e., by a standard uncertainty of 2.9°C (10/2√3) which, when
multiplied by the appropriate coefficient of thermal expansion, gives the
corresponding standard uncertainty u
u
Inthis example δ2> 0 et kuc2-δ2> 0.


Suppose a measuring machine measures the length of 100 mm bars with a standard
uncertainty u
Differential thermal expansion is 22 µm/m/°C - 9 µm/m/°C = 13 µm/m/°C, corresponding to a mean bias of:
13 µm/m/°C x 5 °C x 0,100 m = 6,5 µm
The standard uncertainty associated with temperature variation (modeled as a rectangular distribution) is given by:
utemp = 13 µm/m/°C x 0,100 m x 2,9°C = 3,8 µm
within-laboratory uncertainty: u
Therefore:


![]()
with a 95% level of confidence (K=2)