Statistical Tools for Laboratory Technicians
Statistical tools are widely used in laboratories. They are used on a daily basis to evaluate certain data or simply to report a result based on several measurements.
Some of these tools seem obvious to us because they are, or used to be, part of our academic background. Others may seem more difficult to use due to a lack of familiarity with their purpose and fields of application.
We will attempt to demystify the most commonly used statistical tools through numerical examples.
Preamble:
Tests are not designed to prove or disprove hypotheses.
The hypothesis that we will attempt to reject in the various examples is
generally always the one stating that there is no change, for example, no
difference between two standard deviations within a population. For this reason,
it is called the null hypothesis, H
There is also the alternative hypothesis H
This methodology is based on five distinct steps:
1: Formulate the practical problem in terms of hypotheses.
2: Calculate a statistic (T), a function derived solely from the data. It must have two characteristics:
2.1. It must behave differently when H
2.2. Its probability distribution must be calculable under the assumption
that H
It is also desirable for probability distribution tables to be available.
3: Select a direction. We must be able to determine whether
T tends to support H
One should never conclude by accepting H
4: Determine the significance level α. This is the risk
taken when rejecting hypothesis H
It is also possible not to reject H

5: Determine whether the value of T is close to the non-decision region or whether it falls well within the conditions defined by the distribution tables.
The vast majority of tests assume that the distributions are normal.
Numerical examples may sometimes use errors that are unusually large compared with equipment performance. The primary objective is to illustrate the difficulties involved in decision-making.
![]()
Refresher on the normal distribution:
The normal, or Gaussian, distribution plays an important role in statistics and measurement. Its familiar bell-shaped curve is used to calculate the probability of finding a result within a given range. The x-axis represents the value of the variable being considered, and the y-axis represents the probability density.

The shaded area represents the probability that the value lies between the
bounds x
µ represents the expected value, or mean, and the width of the distribution is given by the variance σ². The square root of the variance corresponds to the standard deviation σ, which characterizes the dispersion of the values around the mean.
Average and standard deviation:
Suppose that n = 10 measurements yield the following values:
| 10,2 |
9,7 |
10,3 |
10,1 |
9,8 |
10,5 |
10,0 |
9,6 |
10,4 |
9,5 |
The mean given by:

x̄ is used as the estimate of the population mean μ and is equal to: 10,01
The standard deviation is given by:

s estimates σ and is equal to: 0,348

By calculating ( f(10.01) ), we obtain 1.146, and ( f(10.5) = 0.425 ), using the Excel formula =NORM.DIST(10.5,10.01,0.348,FALSE) to determine the Gaussian curve.
Now suppose that this series of 10 measurements is repeated a large number of times. The average of all the x values tends toward the population mean, µ.
The population standard deviation of the mean is given by:

Variance equals the standard deviation squared: s²
![]()
One-sample Z-test for a population mean with known variance
Objective: to assess the significance of the difference between the mean µ₀ obtained during a process and the mean of a sample taken from that process for quality control purposes.
Constraints: the process variance σ² must be known, and the distribution must be normal.
Problem: an automatic burette dispenses 400 µL of reagent (µ₀) during a measurement. The standard deviation σ of this process is 10 µL.
Each month, the operator performs 10 measurements of the dispensed volume and obtains a mean x̄ = 406 µL.
What can be said about the dispensing quality of the burette?
Null and alternative hypotheses:
Scenario No. 1 H₀: μ = μ₀ and H₁: μ ≠ μ₀ (Two-tailed hypothesis
test)
Scenario No. 2 H₀: μ = μ₀ and H₁: μ > μ₀ (One-tailed hypothesis test)
Risk level: α = 0.05, corresponding to 0.025 in each tail for a two-tailed test.
The Normal Distribution Critical Values Table For α = 0.05, this gives Z = 1.64 for a one-tailed test and Z = 1.96 for a two-tailed test.
Statistical test:

Z = 1.9

Conclusion:
For scenario 1 (two-tailed distribution), the Z-test statistic falls between
-1.96 and +1.96; therefore, the null hypothesis H
For scenario 2 (one-tailed distribution), the Z-test statistic is greater
than 1.64; therefore, the null hypothesis H
![]()
One-Sample t-Test for a Population Mean with Unknown Variance
Objective: To study the significance of the difference between the mean µ₀ obtained during a process and a reference value.
Constraints: The distribution must be normal, and the sample size must be small (n < 20).
Problem: An automatic burette dispenses a fixed mean volume (µ₀) of 405 µL of reagent over 9 measurements. The estimated standard deviation, s, of this process is 4 µL.
The equipment certificate reports a mean value x̄ = 400 µL.
What can be said about the dispensing quality of the burette?
Hypotheses and alternatives:
Scenario No. 1 — H₀: μ = μ₀ and H₁: μ ≠ μ₀ (Two-sided alternative hypothesis)
Risk level: α = 0.05, corresponding to 0.025 in each tail for a two-tailed test
The Student’s t-distribution critical values table
For α = 0.05 and ν = 9 − 1 = 8 degrees of freedom, the Student’s t critical
value is:t₈;
Statistical test:

t = 3.75

Conclusion:
For the two-tailed distribution scenario, the t-value is greater than 2.3
and the null hypothesis H
![]()
Z-Score
Objective: The Z-score is used to compare a value to a population. If the Z-score is less than or equal to 2.0, corresponding to a probability of approximately 95%, it can be concluded that the method, or the value within the considered range level, is accurate.
Constraints: None.
Problem: Calculation of a laboratory Z-score based on the
10 results obtained during an interlaboratory comparison: x
| 10,2 |
9,7 |
10,3 |
10,1 |
9,8 |
10,5 |
10,0 |
9,6 |
10,4 |
9,5 |
Statistical test:

x
Conclusion:
The z-score is less than 2; the value is satisfactory.
![]()
Fisher F-Test for Comparing Two Variances
Objective: This test is used to study the relative magnitude
of the variances of two series of measurements from two independent processes,
P1 and P2. The larger variance is always placed in the numerator: s
Constraints: It is limited to two series of measurements, which must follow a normal distribution.
Problem: A laboratory has implemented a new piece of equipment,
P2, and hopes that this new process will produce a standard deviation, s
What can be said about the variances of these two populations? Is it likely that they differ?
Hypotheses and alternatives:
Scenario H₀: s₁ = s₂ versus H₁: s₁ > s₂ — right-tailed test
Statistical test:

| N° | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | x̄ | s |
| P1 | 10.2 |
9.7 |
10.3 |
10.1 |
9.8 |
10.5 |
10.0 |
9.6 |
10.4 |
9.5 |
10.01 | 0.348 |
| P2 | 11.3 | 11.1 | 11.5 | 11.5 | 11.0 | 11.4 | 11.3 | / | / | / | 11.30 | 0.191 |
s

F must be compared with the Table de Fisher-Snedecor, which requires knowing the degrees of freedom for each of the two series. In our case, we have 10 and 7 values, respectively, so ν1 = 10 − 1 = 9 and ν2 = 7 − 1 = 6.
The table gives a critical value of Fc = 4.10, which is greater than 3.3. This means that the null hypothesis cannot be rejected. Therefore, the laboratory cannot state, with 95% confidence, that its new equipment provides a smaller standard deviation.
![]()
chi-square test
Objective: To study the difference between the sample variance, s², and the variance of a population, σ².
Constraints: It is assumed that the sample values follow a normal distribution.
Problem: I perform 10 measurements on sample E1, which gives me a standard deviation of s = 3.46.
Regular monitoring of sample E1 over several months gives me a standard deviation of σ = 3.01.
Has the variability of my sample changed?
Hypotheses and alternatives:
Scenario No. 1 H₀: s = σ and H₁: s ≠ σ (Two-tailed hypothesis)
Scenario No. 2 H₀: s = σ and H₁: s > σ (One-tailed hypothesis)
Statistical test:

Let s be the standard deviation of my laboratory, with 9 degrees of freedom (10 − 1), and let σ be the standard deviation of the population.

By consulting the chi-square, we observe that for 9 degrees of freedom and a 95% probability, χ² = [2.70; 19.02] for Scenario 1 and χ² = [3.33; 16.92] for Scenario 2. Since the χ² value falls within the limits, the null hypothesis cannot be rejected. The difference between the two variances is not significant; my sample has not changed..
Critical Values Table for the Normal Distribution
Critical Values Table for Student’s t-Distribution