Selecting Control Limits for Control Charts

Control charts are widely used by laboratories to monitor a characteristic that helps ensure that a test or analysis remains reliable. Walter A. Shewhart introduced the concept of control charts in the 1920s.

Several types of control charts can be developed to monitor individual observations or groups of data. These different charts can help detect changes or trends in the characteristics being monitored.

Control charts for variable data and attribute data include the following:

x̄ and R control chart, where R is the difference between the maximum and minimum values in each data set (range);

x̄ and s control chart, where s is the standard deviation of the data;

Individuals control chart, used when very few measurements are available;

Attribute control charts, for example:
- p-chart: fraction of nonconforming items;
- np-chart: number of nonconforming items;
- c-chart: number of nonconformities;
- u-chart: average number of nonconformities per inspection unit.

For each of these control charts, the applicable distribution assumptions must be satisfied:
The data must follow a normal distribution. If normality is not demonstrated, the overall performance of these charts may be poor and may generate false alarms;
Control limits are calculated from process data. When calculating these limits, as much data as possible should be collected, with at least 25 measurements;
The sample size should be kept as constant as possible.

Let us revisit all these examples using numerical applications:

x̄ and R Control Chart

Suppose a laboratory wants to verify, on each day the equipment is used, that the equipment remains compliant. To do so, it performs (n = 3) measurements X1....X3.

Days
1
2
3
Mean Max - Min = R
1
9.7
10.3
10.1
10.0 0.6
2
9.8
10.2
10.5
10.2 0.7
3
10.2
10.3
10.1
10.2 0.2
4
10.4
10.3
10.0
10.2 0.4
5
9.7
10.5
10.2
10.1 0.8
  10.2 0.54

 

The laboratory determines the limits of its control chart, which are updated dynamically based on the repeated measurement results:

x̄ and s Control Chart

Whenever possible, the sample standard deviation should be used when the sample size (n) is large, for example, when (n > 10).

Days
1
2
3
Mean s
1
9.7
10.3
10.1
10.0 0.31
2
9.8
10.2
10.5
10.2 0.35
3
10.2
10.3
10.1
10.2 0.10
4
10.4
10.3
10.0
10.2 0.21
5
9.7
10.5
10.2
10.1 0.40
  10.2 0.27

 

The laboratory determines the limits of its control chart, which are updated dynamically based on the repeated measurement results:

Individuals Control Chart

There are many practical applications in which a measurement consists of only one observation (n = 1).

This may occur with very slow processes or with processes in which measurement is very costly, such as destructive testing.

An individuals control chart for variable data is appropriate for this type of situation.

The individuals control chart uses the moving range between two successive subgroups or two consecutive days (m=5 for example) to estimate process variability.

The moving range is calculated as follows:

Days
Moving Range = MR
1
9.7
0.0
2
9.8
0.1
3
10.2
0.4
4
10.4
0.2
5
9.7
0.7
m=5 10.0 0.35

 

The laboratory determines the limits of its control chart, which are updated dynamically based on the repeated measurement results.

By default, since only one value is obtained each day, a moving range of two successive values will be used (m = 2).

 

Attribute Control Charts

Attribute control charts are used for discrete or count data.

The characteristic of interest is simply the classification of the observation into a specific category. For example, manufactured products may be measured but classified only as defective/nondefective, conforming/nonconforming, or pass/fail.

Other situations may involve monitoring the number of nonconformities associated with a product in order to detect errors such as an incorrect name, missing or inaccurate information, or unidentified services.

Variable control charts are not suitable for every situation.

When the variable or characteristic of interest is the number of nonconformities per unit, p-charts and np-charts are not appropriate.

The p-Chart: Fraction Nonconforming

The characteristic of interest can be classified into one of two categories: pass/fail, conforming/nonconforming, and so forth.

For simplicity, the term “nonconforming” will be used as a general reference, regardless of the final categories selected.

The following notation is used:
- n: number of products examined (lot size or sample size);
- m: number of subgroups;
- X: number of nonconforming products found in a sample of size (n), where (X \leq n);
- p: probability that the characteristic is nonconforming. This parameter is often unknown and must be estimated;
- p̂: sample fraction nonconforming.

By definition, this probability is given by the following relationship:

Days
n X: Nonconforming
1
100
8 0.08
2
98
7 0.07
3
104
6 0.06
4
99
4 0.04
5
97
5 0.05
m=5  Estimator of n = 99.6 0.06

 

If the lot size is constant—for example, 100—use (n = 100) directly.
If the lot size varies, as in the example above, use the average lot size as the estimate of (n). This approach is simpler than establishing different control limits for each group of lots having the same size.

The lower control limit (LCL) will be set to zero if the calculated value is negative.

The np-Chart: Number of Nonconforming Items

The np-chart is a variation of the p-chart in which the actual number of nonconforming products is plotted.

For the same process, the np-chart and the p-chart provide identical information. In other words, if the p-chart indicates that a process is out of control, the np-chart will also indicate that the same process is out of control.

One reason the np-chart is an attractive alternative to the p-chart is that it is easier to interpret because it displays an absolute number.

For the np-chart, the average fraction nonconforming can be calculated as follows:

Days
n X: Nonconforming
1
100
8 0.08
2
98
7 0.07
3
104
6 0.06
4
99
4 0.04
5
97
5 0.05
m=5  Estimator of n n = 99.6 0.06

 

It can be noted that the control limits of the np-chart are proportional to those of the p-chart, with a factor of n.
The lower control limit (LCL) will be set to zero if the calculated value is negative.

The c-Chart for Monitoring Nonconformities

When the subgroup size n remains constant from one subgroup to another, the c-chart is an appropriate control chart for monitoring nonconformities.
For the c-chart, the following notation is used:

- n: number of units inspected, or sample size, which may be (n = 1) or greater;
- m: number of subgroups;
- X: number of nonconformities per inspected unit or per subgroup;
- c: average number of nonconformities:

Days
n X: Nonconforming
1
100
8
2
100
7
3
100
6
4
100
4
5
100
5
m=5  6

 

The lower control limit (LCL) will be set to zero if the calculated value is negative.

The u-Chart for Monitoring the Average Number of Nonconformities

As with the p- and np-charts, the c- and u-charts provide identical results when the sample size remains constant.

For a u-chart, however, the sample size does not need to remain constant from one subgroup to another.

Let ui be the average number of nonconformities for the (i)th subgroup (i = 1, 2,...., m), where:

Days
n X: Nonconforming ui 
1
100
8 0.08
2
100
7 0.07
3
100
6 0.06
4
100
4 0.04
5
100
5 0.05
m=5  0.06

 

The lower control limit (LCL) will be set to zero if the calculated value is negative.

Limitations of Use

The control charts presented generally apply to processes considered to involve long, continuous runs—for example, when a certified reference material (CRM) is used 100 or 200 times.

These charts are not suitable for short production or inspection runs.

Short-run control charts for attribute data are, in fact, standardized control charts.

The attribute value of the characteristic being monitored is standardized and plotted on a control chart.

To illustrate this, consider the standardized value based on the number of nonconformities—that is, the c chart.

The standardized value, which follows a normal distribution, is:

Under the following conditions:
• Each data point is standardized.
• The standardized random variable Zi follows a standard normal distribution.
• The center line for all standardized charts is zero.
• The control limits for all standardized charts are −3 and +3.

 

Days
n X: Nonconforming Zi 
1
100
8 0.81
2
100
7 0.40
3
100
6 0.00
4
100
4 -0.81
5
100
5 -0.40
m=5  6

 

 

The process is under control


Burr Constants