Calibration / Verification Interval Calculator

Tool based on an AHP risk analysis.

The user only assigns the equipment ratings and must evaluate each criterion from the equipment user’s perspective.

The overall score provides a recommended interval.

Equipment Rating
Criterion AHP
Weight
Rating
Measurement Criticality 49.6 %
Metrological History 24.5 %
Operating Conditions 12.3 %
Internal Control 8.5 %
Personnel Competency 5.0 %
Principle: the higher the rating, the more favorable the conditions are for extending the interval. The overall score is the sum of each AHP weight multiplied by the corresponding criterion rating. All pairwise matrix selections have been justified.
Additional Information
Validation: The matrix incorporates requirements from certain standards that limit calibration or verification intervals to no more than one year. For the equipment concerned, consistent ratings therefore result in intervals that comply with the applicable standards. Matrix consistency has also been statistically verified and demonstrated.


Result
1.00
Overall Weighted Score
High Risk
1 Year
Rating Guide
1 — High Risk: unfavorable conditions, significant drift or uncertainty, critical use, or insufficient control.
3 — Moderate: acceptable conditions requiring regular monitoring.
5 — Highly Favorable: controlled conditions, stable history, low criticality, or robust internal control.
Proposed Decision Rules
Score < 2.60
1 Year
≥ 2.60 to < 3.30
2 Years
≥ 3.30 to < 4.00
3 Years
≥ 4.00
5 Years
These thresholds may be adapted to your internal rules, applicable requirements, and metrological experience. However, they are an integral part of the method validation and were established on a statistical basis. Any change should therefore be made cautiously because it could produce an interval that no longer complies with certain applicable standards or regulatory requirements.