Example of Measurement Uncertainty Estimation in Accordance with ISO/IEC Guide 98-3
The purpose of this exercise is to estimate the calibration uncertainty of a 1 Ω resistor.
5M Method:

Data :
A resistor is calibrated using a reference resistor. Both resistors are connected in series to a direct-current source, so the same current flows through each resistor.
A digital voltmeter is used to measure the voltage drop across each resistor: resolution = 0.1 × 10⁻⁶ V
Using a bypass switch, the voltage across each resistor is measured in turn.
The resistance being calibrated is denoted R
The corresponding voltage drops are denoted V
The value of the reference resistor and its expanded uncertainty, with (k
= 2), are obtained from the calibration certificate: R
The uncertainty due to stability is: 5.10
Ten voltage measurements are taken for each resistor.
Preliminary statement:
Since the same current flows through both resistors, the following relationship can be written:

The factors contributing to the measurement uncertainty are:
The uncertainty associated with the repeated measurements of V
The uncertainty associated with the repeated measurements of V
The uncertainty associated with the digital voltmeter resolution;
The manufacturer’s accuracy specification for the voltmeter, equal to (0.025%
+ 2) digits of the measured value.
The two digits are considered to include the reading-resolution contribution;
The uncertainty associated with the certified value of the reference resistor;
The uncertainty associated with the stability of the reference resistor value;
The uncertainty associated with the difference between the laboratory temperature
and the temperature at which the reference resistor was calibrated.
The uncertainty associated with R

Measurement Results:
| N° |
V |
V |
| 1 | 99,91963 | 99,92478 |
| 2 |
99,91970 | 99,92474 |
| 3 |
99,91968 | 99,92470 |
| 4 | 99,91977 | 99,92492 |
| 5 | 99,91986 | 99,92488 |
| 6 | 99,92058 | 99,92561 |
| 7 | 99,92055 | 99,92558 |
| 8 | 99,92008 | 99,92506 |
| 9 | 99,91987 | 99,92479 |
| 10 | 99,91968 | 99,92468 |
| Mean | 99,919940 | 99,924974 |
| Standard deviation | 3,544.10 |
3,461.10 |
The Type A uncertainties associated with repeatability, expressed in mV, are given by the following equations:

The number of degrees of freedom is 9:(n-1=10-1)
Resistance Calculation R

Determination of the Uncertainty in R
We will take the partial derivatives of the function defining
R

The uncertainty associated with the voltmeter resolution is included twice, once for each resistor. It follows a rectangular distribution and, for a voltage of approximately 100 mV, is equal to:

The uncertainty associated with the reference resistor is assumed to follow a normal distribution. Since the stated uncertainty is an expanded uncertainty with k=2, it must be divided by 2 to obtain the standard uncertainty used in the calculation:

The uncertainty associated with stability follows a rectangular distribution and is equal to:

The uncertainty associated with temperature is considered negligible because the calibration was performed under the temperature conditions specified in the reference resistor calibration certificate, namely 20 °C.
Using the root-sum-of-squares of the various uncertainty components, we obtain:

Using a coverage factor of k = 2, UR
Since only 10 measurements were performed, we must verify whether a coverage factor of k=2 can be used by applying the Welch–Satterthwaite formula:

The effective number of degrees of freedom is sufficiently
large; therefore, the resistance value can be expressed as follows: R
Effect of Correlation Between the V
Indeed, the same voltmeter was used for both sets of measurements,
which were performed during the same measurement session. Therefore, the variables
V
We must therefore determine the correlation coefficient of the data ((see the Error Propagation Example) in order to apply the law of propagation of uncertainty, including the correlation term.
We need to calculate the following differences:
| N° |
|
||
| 1 | 99,91963 - 99,919940 | 99,92478 - 99,924974 | 6,014.10 |
| 2 |
99,91970 - 99,919940 | 99,92474 - 99,924974 | 5,616.10 |
| 3 |
99,91968 - 99,919940 | 99,92470 - 99,924974 | 7,124.10 |
| 4 | 99,91977 - 99,919940 | 99,92492 - 99,924974 | 9,18.10 |
| 5 | 99,91986 - 99,919940 | 99,92488 - 99,924974 | 7,52.10 |
| 6 | 99,92058 - 99,919940 | 99,92561 - 99,924974 | 40,704.10 |
| 7 | 99,92055 - 99,919940 | 99,92558 - 99,924974 | 36,966.10 |
| 8 | 99,92008 - 99,919940 | 99,92506 - 99,924974 | 1,204.10 |
| 9 | 99,91987 - 99,919940 | 99,92479 - 99,924974 | 1,288.10 |
| 10 | 99,91968 - 99,919940 | 99,92468 - 99,924974 | 7,644.10 |
| Sum | / | / | 108,23.10 |
| n(n-1) | / | / | 90 |
Sum /n(n-1) |
/ | / | 1,202.10 |
Let us calculate the correlation coefficient:

r = 0,109 (The correlation coefficient is closer to 0 than to 1, indicating that the data are only weakly correlated).
Continuing the estimation of the second-order term gives:

A negligible quantity that must be added to the initially estimated uncertainty:

Using a coverage factor of k = 2, UR
The resistance value can therefore be expressed as follows:
R