DemarcheISO17025
  • Home
  • PREvious
Propagation of Errors

Example of Error Propagation

The law of propagation of errors is particularly useful when, for example, limited resources require you to use the same measuring instrument to measure different dimensions of the same object, which are then used to calculate its surface area.

Degrees of Freedom

Degrees of Freedom

This is an aspect that is often overlooked when evaluating measurement uncertainty. However, the coverage factor can significantly affect the final expanded uncertainty.

The GUM (Guide to the Expression of Uncertainty in Measurement)

Fundamental Metrology Concepts

- Measurement process modeling: Expressing mathematically how all the information available to the experimenter is used, which can sometimes be difficult or even impossible;

- Measurement uncertainty evaluation: Estimating the uncertainty of the output quantity based on the uncertainties associated with the input quantities;

- Identification of uncertainty sources: The need to identify all possible sources of error, including influence quantities, calibration errors, reading errors, etc.;

- Practical implementation: Applying this approach can be particularly challenging in the case of complex testing procedures.

General Approach:

Analysis of the measurement process to establish a mathematical relationship incorporating all the information available to the user:

Y (measurand) = f(X1, X2, ……Xn)

- Identification of the components and sources of measurement errors.
- Evaluation of the standard uncertainties associated with each individual source contributing to the uncertainty of the measurement result, for each input quantity.
- Determination of the combined standard uncertainty (law of propagation of uncertainty).
- Determination of the expanded uncertainty and reporting of the measurement result.

The 5M Method (Ishikawa Method)

The 5M method is used to identify potential sources of error by analyzing the measurement process and listing all components that contribute to the uncertainty of the measurement result.

Each uncertainty component is evaluated separately using Type A or Type B evaluation methods;

Root Mean Square (RMS)

The combined standard uncertainty is then calculated by combining the individual uncertainties according to the law of propagation of uncertainty, based on the mathematical model (f) used to calculate the measurement result.

he combined standard uncertainty, u(y), is the square root of the sum of the squared partial derivatives, each multiplied by the square of its associated standard uncertainty (assuming uncorrelated input quantities).

Some useful formulas for calculating partial derivatives.

The root-sum-of-squares combination of individual contributions (which is also characteristic of the classical theory of error propagation) takes into account the fact that the probability of all individual components simultaneously reaching their maximum values and having the same sign is very low.

The root-sum-of-squares method is based on the Pythagorean theorem:

However, this does not preclude the use of conservative approaches in certain cases, where all or some of the uncertainty components are combined by simple arithmetic addition.

It should be noted that such an estimate may be highly approximate when the test does not consist of a straightforward sequence of operations, each characterized by individual uncertainty contributions \(U_i\) (as in the case of a series of weighings or volume measurements combined with instrument readings), but instead involves a single operation or a set of simultaneous operations in which several sources of uncertainty act concurrently.

In such cases, individual uncertainty components may not propagate directly to the final result. Their effects may instead be attenuated or amplified, depending on the measurement model and the associated sensitivity coefficients.

Note: Partial derivatives are referred to as sensitivity coefficients.

It is virtually impossible to evaluate individually the effect of every influence quantity on measurement uncertainty.

Furthermore, experience has shown that uncertainties determined using the analytical-mathematical approach are generally lower.

For well-established test methods described in standards or other normative documents, where the fundamental performance characteristics are known, the validation of the method is considered to have been addressed through the standardization process.

However, the laboratory must still demonstrate:
1. Strict compliance with the standard and its associated work instructions;
2. Knowledge and control of the relevant influence quantities;
3. The ability to provide, upon customer request, a consistent and technically justified estimate of measurement uncertainty associated with the test result.

This demonstration should be based on:
- Primarily, repeatability studies using reference materials;
- Reproducibility studies, where applicable;
- Direct comparisons of results to demonstrate that the testing process is adequately controlled.

In addition to the complete testing procedure, the laboratory should document the evaluation of measurement uncertainty.

The combined standard uncertainty will generally be reported as an expanded uncertainty (k = 2 — although this is not always appropriate). The uncertainty statement shall clearly indicate:
- That the expanded uncertainty is based on an assumed normal distribution and corresponds to a coverage probability of approximately 95%.

This means that, under the assumptions of the uncertainty model, the stated interval is expected to contain the value of the measurand with a coverage probability of approximately 95%.
- Any significant influence quantities that have been excluded from the uncertainty evaluation, such as sampling.

In certain cases, additional precautions may be necessary, and a coverage factor of k = 3 may be adopted, corresponding to a coverage probability of approximately 99.7% under the assumption of a normal distribution.

Consequently, the probability that the value of the measurand lies outside the stated expanded uncertainty interval is very low.

Technical note: For a normal distribution, k = 2 corresponds to approximately 95.45% coverage, while k = 3 corresponds to approximately 99.73%.

These coverage probabilities are not automatically applicable to every uncertainty evaluation; the effective degrees of freedom and the assumed probability distribution must also be considered.

Furthermore, an expanded uncertainty interval is not, strictly speaking, a classical statistical confidence interval.

The Law of Propagation of Uncertainty

When the input quantities are independent (uncorrelated), the following equation can be used:

However, when the uncertainty associated with one input quantity is correlated with that of another input quantity (for example, when different dimensions are measured using the same caliper), the following formula must be used to account for correlations in the propagation of uncertainty:

The correlation coefficient r(xi,xj) characterizes the degree of correlation between the input quantities:
- For uncorrelated data, r = 0.
- For perfectly correlated data, r = -1 or +1.
- For partially correlated data, r ranges between -1 and +1.
To estimate this correlation coefficient, the covariance between the input quantities must first be evaluated.
The covariance is estimated using the following formula:

The correlation coefficient is given by the following ratio:

Degrees of Freedom

The combined standard uncertainty Uc(y) may not be sufficiently reliable if some of the dominant standard uncertainty components used in its calculation are derived from:
- A small sample size;
- A reasonable estimate based on expert judgment;
- An average calculated from different values whose variability directly affects the measurement result.
To quantify the reliability of Uc(y), the effective number of degrees of freedom is calculated using the Welch–Satterthwaite equation:

The reliability of the combined standard uncertainty Uc(y) increases as the effective number of degrees of freedom, Veff, increases.
For Type A uncertainty evaluations based on a series of repeated measurements, the number of degrees of freedom is equal to the number of measurements minus one (n-1).
When uncertainty is evaluated using simple linear regression with two fitted parameters (slope and intercept), the number of degrees of freedom is (n-2).

Copyright © DemarcheISO17025.com