Interlaboratory Comparisons

Types of Interlaboratory Comparisons

These comparisons are essential for ensuring the reliability of test results and, above all, for obtaining materials or liquid solutions with established properties for the parameters of interest.

Definitions

Definitions Related to Proficiency Testing

Terminology relating to proficiency testing is important to avoid confusing, for example, a reference laboratory with a participating laboratory.

7.7 - Ensuring the Validity of Results

7.7.1 The laboratory shall have a procedure for monitoring the validity of results. The resulting data shall be recorded in such a way that trends are detectable and, where practicable, statistical techniques shall be applied to review the results. This monitoring shall be planned and reviewed and shall include, where appropriate, but not be limited to:
a) regular use of reference materials or quality control materials;
b) regular use of alternative instruments that have been calibrated to provide traceable results;
c) functional checks of measuring and testing equipment;
d) use of check or working standards with control charts, where applicable;
e) periodic intermediate checks of measuring equipment;
f) replicate tests or calibrations using the same or different methods;
g) retesting or recalibration of retained items;
h) correlation of results for different characteristics of an item;
i) review of reported results by competent laboratory personnel;
j) intralaboratory comparisons;
k) blind tests.

7.7.2 The laboratory shall monitor its performance by comparison with the results of other laboratories, where available and appropriate. This monitoring shall be planned and reviewed and shall include, but not be limited to, either or both of the following:
a) participation in proficiency testing;
b) participation in interlaboratory comparisons other than proficiency testing.

Definition of proficiency testing: This is a means of comparing a laboratory's results with those of other laboratories. It is widely used by most laboratories and offers the following benefits:
• Periodic, independent, objective assessment of the quality of routine analyses performed by the conformity assessment body (CAB);
• Feedback that can lead to technical improvements, correction of instrument drift or deficiencies in technical skills, and completion of the qualification of new personnel;
• Sharing information on the performance of methods and instruments (generally underused: ISO/IEC 17043-accredited providers are not required to report findings to standards committees; moreover, techniques used to obtain the same result can differ substantially);
• Assessment of analytical quality in a specific field (feedback that could be provided if, for example, following a change to a standardized test method, the proficiency testing provider observed that all participants in a given industry obtained incorrect or widely scattered results, suggesting that the method is subject to significant interference or is unreliable).

Ideally, proficiency testing samples should be similar to the actual samples tested by the CAB and sufficiently stable and homogeneous not to affect assessment of the participants' performance. Otherwise, sample heterogeneity may outweigh between-laboratory variability, undermining the purpose of the comparison.

Definition of an interlaboratory comparison:

Its scope and objectives are broader: certification of a material to assign a reference value, validation of a test method, and evaluation of the method's overall performance. Generally, organizations conducting this type of study invite nationally or internationally recognized laboratories to reduce the variability of the results.

Proficiency testing is a form of interlaboratory comparison, but it represents only one of the possible purposes of such comparisons.

7.7.3 Data from monitoring activities shall be analyzed and used to control and, where applicable, improve laboratory activities.

If the results of monitoring data analysis fall outside predefined criteria, planned action shall be taken to correct the problem and prevent incorrect results from being reported.

In general, a control chart is a visual tool for quickly assessing trends in a monitored parameter, provided that the reference material or solution (certified or not) can be monitored over time and enough data are available to identify a trend.

A good approach is to use Excel trendlines. Choose simple models and establish monitoring limits, which may be tighter than those of the primary or secondary standard.

Methodology:

a) regular use of reference materials or quality control materials:

Some definitions:

Reference material (RM): a material sufficiently homogeneous and stable with respect to specified properties, which has been established to be fit for its intended use in measurement or in the examination of qualitative properties.
RMs are generally prepared in batches for which property values are determined, within stated uncertainty limits, from measurements on a sample representative of the entire batch. (VIM)

Certified reference material (CRM): reference material accompanied by documentation issued by an authoritative body and providing one or more specified property values, with associated uncertainties and traceabilities, using valid procedures. (VIM)

When properly used, they provide particularly valuable information about confidence in the quality of measurement results.
According to ISO Guide 33, “Reference materials — Good practice in using reference materials,” they are used for instrument calibration, method validation, assessment of methods and performance, metrological traceability of measurement results, and evaluation of their uncertainty.

Some confusion commonly arises from the indiscriminate use of terminology that does not always reflect the intended application. Calibration standards, CRMs, RMs, quality control samples, reference standards, standard reference materials, and similar terms are often used interchangeably or with very similar meanings.

RM producers may claim that a material is a CRM even though not all the requirements of the definition above have been met.

Sometimes a technician cannot purchase an appropriate RM or CRM because none is commercially available.

A special case is so-called matrix reference materials. They contain more than one element, each with a characterized nominal value and an associated confidence interval, generally based on the dispersion of results from participating laboratories. Such materials are commonly produced through interlaboratory comparisons using several analytical techniques. Some elemental values may also be provided for information only when no consensus could be established from the results.

For multielement liquid solutions, it is easier to prepare them from certified single-element solutions and thus obtain a certified final solution. Very few laboratories in France are accredited to produce such solutions, and the accredited scope is generally limited to relatively basic materials, such as pH, conductivity, and single-element solutions.

It is worth recalling that a certificate providing appropriate quality assurance should bear the logo of the relevant accredited body and refer to ISO Guide 34 or ISO 17034 (the content of the certificate is defined in ISO Guide 31).

However, before selecting and using any reference material in the laboratory, the user is responsible for understanding its characteristics, benefits, and limitations. In some cases, properties are guaranteed only under specific conditions.

The user must closely follow the recommendations for material storage and comply with the shelf life stated on the certificate.

b) regular use of alternative instruments that have been calibrated to provide traceable results:

This useful approach requires several pieces of equipment based on identical or different physical principles that can yield comparable results (in the same units). For example, copper concentration in an aqueous solution may be determined by ICP spectrometry, titrimetry, and photometry (with standardized methods for all three techniques), with results expressed as a percentage or in ppm.

This assumes that the laboratory has several identical or different measurement systems. Small laboratories may therefore find it difficult to implement. The CAB also needs sufficient analytical experience to account for interferences that may arise with certain techniques that are less suited to the analysis in question. A standardized method does not necessarily work for every matrix or for all interfering contaminants that may increase the measurement error.

This approach is particularly useful for validating an in-house, nonstandard method. However, substantial work is required to demonstrate that the results are not influenced by the particular limitations or unpredictable effects of the different analytical techniques used.